The DLM-080 FUTURE Digital Recorder – User’s Manual
26.11.2004
17
6. ALARMS
The DLM-080 Recorder enables setting of alarm thresholds in relation to set value. It is
required to preset the alarm threshold value in the F70-F77 (SETUP) functions. In order
to activate alarm press the
key in field of the channel, for which exceeded
threshold value is to be signalled. In case if value = 0 is set in the F70-F77 functions,
threshold exceeding control is off, and alarm is off no matter what is current state of the
key.
Blinking diodes in the keys (for proper channels) and intermittent signal on the
RE9 relay output indicate occurrence of an alarm.
Alarm enable unit, which is preset in SETUP (F70-F77 cells), depends on measurement
resolution !
E.g.: Channel 4 is defined as temperature measurement carried out with the Pt-100 sensor.
Measurement resolution (accuracy) is 0.1°C. Set value: 27 °C. In order to activate alarm
in the F73 cell value 5 has been entered. Then alarm will be activated as soon as read out
temperature value exceeds 27.5°C.
7. CO-OPERATION OF THE RECORDER WITH MASTER COMPUTER (RS-
485)
In order to connect the Recorder to master computer (RS-485) it is necessary to make
suitable connections between the Recorder and master computer according to Drawing
No. 1, and then to install the DLM-SIMPLE communication application. Each DLM-080
recorder connected to master computer with the RS-485 network shall have attributed its
own number in the RS-485 network (0-31). Specify this number in the F00 (SETUP)
function.
In case, if there are two or more recorders in the network, which have the same
network number attributed, communication with these recorders will be impossible !
8. PRINTING OF RECORDS
Printing data transmission to printer is carried out through the RS-232 serial connector,
with the following transmission parameters :
transmission speed: 9600 BAUD
no parity control
8 data bits